Figure 2.7
Exploded View
of TMA Stage Assembly
10. Attach the thermocouple to the stage as
follows:
a. Position the tip of the thermocouple so
that it bends at a 90°
angle and lies flat against
the platform. It should be
close to, but not touching
the sample.
b. Hold the thermocouple against the stage
assembly and put on the spring clip to
hold the thermocouple in place (see
Figure 2.8).
Installing the Instrument