Running Experiments
TA I
NSTRUMENTS
TMA 2940
3–8
When the sample thermocouple is changed
When the TMA is serviced or repaired
Periodically (approximately once a month)
If you are changing the temperature range of
interest
If the run data obtained seems to be
inaccurate.
Regardless of the size or shape of the sample that you
are running on the TMA, position the tip of the thermo-
couple so that it bends at a 90° angle and lies flat
against the platform. It should be close to, but not
touching the sample.