TA I
NSTRUMENTS
TMA 2940
5–7
Status Codes
Status Codes
Status codes are character strings that are
continuously displayed at the top left of the
TMA instrument display. These codes tell you
what segment is currently being performed by
the instrument.
Table 5.1
Method Status Codes
Code
Meaning
Air Cool
The air cool solenoid is
opened to allow cooling
air flow to the furnace.
Balance
Force is being applied
to balance the system at
the center of the displace-
ment range. This is used
for force calibration.
Calib
The instrument is per-
forming LVDT calib-
ration or the instrument is
in a calibration mode. No
method is running.
Cold
The instrument heater
cannot supply heat fast
enough to keep up with
the thermal program.
This may be caused by a
large ballistic jump in the
program, a faulty heater, a
faulty control thermo-
couple signal.
(table continued)
Summary of Contents for TMA 2940
Page 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Page 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Page 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Page 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Page 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Page 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Page 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Page 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Page 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Page 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Page 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...