The parallel plate rheometer employs cylindrical
samples that are 9.53 mm (0.375 inch) in
diameter and 0.25 to 1.00 mm (0.01 to 0.04
inch) in thickness. Three types of samples can
be used with the rheometer accessory:
•
Moldable powders
•
Resinous materials
•
Sheet materials.
In order to obtain a sample having the correct
cylindrical dimensions, the material must be
prepared according to the directions in this
section. Moldable powders and resinous
materials employ the use of the Sample Encap-
sulating Press (PN 900680.902) and a special set
of dies, the pellet press die set.
To prepare samples follow the instructions
found in these sections:
•
Sample Encapsulating Press
__
see below
•
Preparing Moldable Powders
__
page 4-35
•
Preparing Resinous Materials
__
page 4-38
•
Preparing Sheet Materials
__
page 4-40.
Sample
Encapsulating
Press
This accessory (see Figure 4.11) is used to form
samples into a compact, cylindrical disc. In
order to use it for parallel plate rheometer
samples, you must first remove the Encapsulat-
ing Press hermetic sealing or non-hermetic
crimping die assembly. After these dies have
been removed, the pellet press die set can be
installed.
Parallel Plate Rheometer Accessory
Summary of Contents for TMA 2940
Page 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Page 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Page 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Page 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Page 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Page 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Page 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Page 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Page 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Page 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Page 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...