TA I
NSTRUMENTS
TMA 2940
3–19
Because of its conductivity, helium is the
purge gas recommended when performing
TMA measurements at temperatures below
ambient. Helium reduces the time required
to cool the sample, provides a more stable
sample environment, and reduces cold
drafts on the system mechanism.
Use the following helium purge rates for
subambient operation:
100 mL/min at -100
o
C and above
200 mL/min at -150
o
C and above.
If a denser purge gas than helium is used for
subambient experiments, it is recommended
that the silicon washer found in the
accessory kit be attached to the probe.
Raise the probe to the fully open position,
then place the washer so it is just below the
opening in the stage. This will minimize
cold drafts into the lower balance enclosure.
Let the purge gas flow for 3 to 5 minutes
before adding coolant or starting the run;
this allows a thorough exchange of sample
atmosphere.
Running a TMA Experiment
Summary of Contents for TMA 2940
Page 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Page 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Page 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Page 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Page 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Page 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Page 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Page 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Page 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Page 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Page 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...