Installation
TA I
NSTRUMENTS
TMA 2940
2–10
If the instrument is intact but parts are missing,
contact TA Instruments.
A list of TA Instruments offices can be found in
Appendix A of this manual.
Choosing
a Location
Because of the sensitivity of TMA experiments,
it is important to choose a location for the
instrument using the following guidelines:
In . . .
a temperature-controlled area.
. . .
a clean environment.
. . .
an area with ample working and
ventilation space around the instru-
ment. (Refer to the technical speci-
fications in Chapter 1 for the instru-
ment’s dimensions.)
On . . .
a stable, vibration-free work surface.
Near . . . a power outlet (115 Vac, 50 or 60 Hz,
10 amps). A step up/down line trans-
former may be required if the unit is
operated from a higher or lower line
voltage.
. . .
your TA Instruments thermal analysis
controller.
. . .
sources of compressed lab air and
purge gas supply for use during
cooling and subambient experiments.
Away
from. . .
dusty environments.
. . .
exposure to direct sunlight.
. . .
direct air drafts (fans, room air ducts).
. . .
poorly ventilated areas.
Summary of Contents for TMA 2940
Page 18: ...TA INSTRUMENTS TMA 2940 XVIII...
Page 20: ...Introduction 1 2 TA INSTRUMENTS TMA 2940...
Page 34: ...Introduction 1 16 TA INSTRUMENTS TMA 2940...
Page 36: ...Installation TA INSTRUMENTS TMA 2940 2 2...
Page 62: ...Installation TA INSTRUMENTS TMA 2940 2 28 Figure 2 17 Isolation Transformer Cable Connections...
Page 66: ...Installation TA INSTRUMENTS TMA 2940 2 32...
Page 138: ...Using Your Options TA INSTRUMENTS TMA 2940 4 48...
Page 140: ...5 2 TA INSTRUMENTS TMA 2940 Technical Reference...
Page 150: ...5 12 TA INSTRUMENTS TMA 2940 Technical Reference...
Page 152: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 2...
Page 168: ...Maintenance and Diagnostics TA INSTRUMENTS TMA 2940 6 18...