TA I
NSTRUMENTS
TMA 2940
4–27
Figure 4.9
Dilatometer Sample
Preparation
7. Raise the furnace and rotate it clockwise to
move it off to the side.
8. Remove any previously run samples from
the stage and ensure that no residue remains.
9. Open the probe.
10. Place the vial on the stage.
11. Close the probe until the probe tip is posi-
tioned above the filling medium.
12. Tap lightly on the upper probe assembly, the
probe should oscillate freely within the vial.
13. Close the probe until it just rests on top of
the filling medium. Gently tap the probe to
seat it (see Figure 4.10).
Dilatometer Accessory