12
FIRE/EXPLOSION WARNINGS
1.
DO NOT use this device in the presence of flammable anaesthetics,
other flammable gases or objects, near flammable fluids such as skin
prepping agents and tinctures, or oxidizing agents. Observe appropriate
fire precautions at all times.
2.
DO NOT use this device in oxygen enriched atmospheres, nitrous oxide
(N2O) atmospheres, or in the presence of other oxidizing agents , to
prevent risk of explosion. Ensure that oxygen connections are not
leaking.
3.
Electrosurgical components, such as the Probe, may remain hot
following activation. Keep all electrosurgical equipment away from
flammable materials to avoid combustion.
4.
To prevent the risk of fire, DO NOT replace Generator fuses. If it is
suspected that fuses are damaged, return Generator to Stryker for repair.
PRIOR TO SURGERY
1.
The operator of the SERFAS Energy System should be a qualified
physician, having complete knowledge of the use of this equipment and
awareness of the risks associated with surgical procedures.
2.
The operator of the SERFAS Energy System should have current
knowledge of technological advances in surgical products and
techniques.
3.
The operator of the SERFAS Energy System should read this manual
thoroughly and be familiar with its contents prior to operating the
equipment.
4.
The operator of the SERFAS Energy System should be sure that the
system functions as outlined in this manual prior to a surgical
procedure. The SERFAS Energy System was fully tested at the factory
before shipment.
5.
SERFAS Energy components are designed to be used together as a
system. Use only the appropriate SERFAS Energy Footswitch, Probe,
and RF Generator.
6.
The SERFAS and SERFAS Energy systems are independent systems.
Their components are not compatible with each other.
7.
Operators of the SERFAS Energy System should be experienced in
arthroscopic and electrosurgical practices and techniques.
8.
Carefully unpack the unit and ensure that all components are accounted
for and remain undamaged from shipment. Inspect the probe cable for
any damage to insulation. If damage to any component is detected, refer
to the ‘Service and Claims’ section of this manual.
9.
DO NOT use the SERFAS Energy System on patients with cardiac
pacemakers or other electronic device implants. Doing so could lead to
electromagnetic interference and possible death.
Summary of Contents for SERFAS
Page 2: ......
Page 10: ...8 ...
Page 11: ...SERFASEnergy System Operation and Maintenance Manual Endoscopy Energy System ...
Page 43: ...SERFASEnergy System Bedienungs und Wartungshandbuch Endoscopy Energy System ...
Page 75: ...73 ...
Page 77: ...SERFASSistema Energy Manual de utilización y mantenimiento Endoscopy Sistema Energy ...
Page 111: ...SERFASSystème Energy Manuel d utilisation et de entretien Endoscopy Système Energy ...
Page 143: ...141 ...
Page 145: ...SERFASSISTEMA Energy Manuale di funzionamento e manutenzione Endoscopy SISTEMA Energy ...
Page 179: ...SERFASSistema Energy Manual de Operação e Manutenção Endoscopy Sistema Energy ...
Page 211: ...209 ...
Page 213: ...SERFASEnergy systeem Gebruiks en onderhoudshandleiding Endoscopy Energy systeem ...
Page 247: ...SERFASEnergy system Bruger og vedligeholdelsesvejledning Endoscopy Energy system ...
Page 279: ...SERFASEnergy järjestelmä Käyttö ja huolto opas Endoscopy Energy järjestelmä ...
Page 311: ...SERFASEnergy System Manual för drift och underhåll Endoscopy Energy System ...
Page 343: ...SERFASȈȪıIJȘµĮ Energy ǼȖȤİȚȡȓįȚȠ ȜİȚIJȠȣȡȖȓĮȢ țĮȚ ıȣȞIJȒȡȘıȘȢ Endoscopy ȈȪıIJȘµĮ Energy ...
Page 377: ...SERFASEnergy System Podręcznik użytkowania i obsługi Endoscopy Energy System ...
Page 409: ...407 ...
Page 411: ...SERFAS Energy ΏΑΞθ ե ྶ Endoscopy Energy ΏΑΞθ ...
Page 443: ...SERFAS øœµÕŠ øœµÕŠ Ϣ㓈ᡸ ݠ Endoscopy 能量系统 ...
Page 467: ...SERFAS ø Šþ ˆ Ð Ý ø Šþ ˆ Ð Ý 㧧 G G㥶㰖ὖⰂG ⳛ GGG Endoscopy 櫖嘎滆 柢枪癢 ...
Page 493: ...491 ...