LIFT POTENTIOMETER “BURN-IN” PROCEDURE
1. Unplug the bed power cord from the wall socket.
2. On the footboard control panel, hold down the Bed Motion Lock, Gatch Lock, and Reverse Trend buttons
simultaneously. While still holding down the buttons, plug the power cord into the wall socket. Release the
footboard button. Scroll down to the “Burn-In” menu item and press enter. Then select the “Burn-In Height” item.
Based on the type of bed, a message will be displayed. This message will indicate to the user whether to run the
bed all the way down or all the way up.
3. Hold down the Bed Motion Lock button until the light flashes.
4. Release the button and unplug the power cord from the wall socket.
5. Plug the power cord back in to the wall socket. Run the bed to full up, then full down to verify the bed limits.
6. The distance between the floor and the top of the litter seat section on a standard Secure bed (without a mattress)
should be 16” - 16 3/8” with the litter fully down and 29 1/4” - 29 7/8” with the litter fully up.
WARNING
Ensure there is no patient on the bed while the burn-in procedure is being performed.
Service Information
Summary of Contents for 3002
Page 2: ......
Page 82: ...Return To Table of Contents 82 3006 409 006 REV C www stryker com Bed Assembly Standard Bed ...
Page 83: ...Return To Table of Contents www stryker com 3006 409 006 REV C 83 Bed Assembly Standard Bed ...
Page 84: ...Return To Table of Contents 84 3006 409 006 REV C www stryker com Bed Assembly Standard Bed ...
Page 85: ...Return To Table of Contents www stryker com 3006 409 006 REV C 85 Bed Assembly Standard Bed ...
Page 91: ...Return To Table of Contents www stryker com 3006 409 006 REV C 91 Base Assembly Standard Bed ...
Page 92: ...Return To Table of Contents 92 3006 409 006 REV C www stryker com Base Assembly Standard Bed ...
Page 95: ...Return To Table of Contents www stryker com 3006 409 006 REV C 95 Base Assembly Standard Bed ...
Page 100: ...Return To Table of Contents 100 3006 409 006 REV C www stryker com Lift Assembly ...
Page 134: ...Return To Table of Contents 134 3006 409 006 REV C www stryker com Mechanical Litter Assembly ...
Page 135: ...Return To Table of Contents www stryker com 3006 409 006 REV C 135 Mechanical Litter Assembly ...
Page 164: ...Return To Table of Contents 164 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Page 165: ...Return To Table of Contents www stryker com 3006 409 006 REV C 165 Head End Siderail Assembly ...
Page 166: ...Return To Table of Contents 166 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Page 167: ...Return To Table of Contents www stryker com 3006 409 006 REV C 167 Head End Siderail Assembly ...
Page 168: ...Return To Table of Contents 168 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Page 189: ...Return To Table of Contents www stryker com 3006 409 006 REV C 189 Foot End Siderail Assembly ...
Page 190: ...Return To Table of Contents 190 3006 409 006 REV C www stryker com Foot End Siderail Assembly ...
Page 191: ...Return To Table of Contents www stryker com 3006 409 006 REV C 191 Foot End Siderail Assembly ...
Page 192: ...Return To Table of Contents 192 3006 409 006 REV C www stryker com Foot End Siderail Assembly ...
Page 193: ...Return To Table of Contents www stryker com 3006 409 006 REV C 193 Foot End Siderail Assembly ...
Page 200: ...Return To Table of Contents 200 3006 409 006 REV C www stryker com Notes ...
Page 244: ......