FOWLER POTENTIOMETER REPLACEMENT
TOOLS REQUIRED:
• T27 Torx.
• Wire Cutters.
• 1/2” Open End Wrench.
• 7/64” Allen Wrench.
PROCEDURE
1. Unplug the bed power cord from the wall socket.
2. Manually raise the fowler to 60 degrees.
3. Using a 7/64” Allen wrench, loosen the screw holding the linkage to the potentiometer shaft and remove the linkage
from the shaft.
4. Using a ½” open end wrench, remove the nut holding the potentiometer to the frame and disconnect from the 3
pin quick disconnect.
5. Install the fowler potentiometer to the Fowler platform arm using a ½” combination wrench, making sure to align
the potentiometer alignment tab up with the notch in the fowler frame.
6. Using a multi-meter and small flat blade screwdriver, set the fowler potentiometer to 500 +/-50 ohms, using the
black and white wires from the fowler potentiometer.
7. Using a 7/64” Allen wrench, tighten the set screw on the fowler potentiometer linkage attaching to the potentiometer.
Make sure to double check the ohms reading after tightening the set screw; readjustment maybe necessary.
8. Refer to the Fowler Potentiometer “burn-in” procedure to set the Fowler angle.
FOWLER POTENTIOMETER “BURN-IN” PROCEDURE
Note:
It requires
two people
to enable the diagnostics mode for the bed.
PROCEDURE
1. Unplug the bed power cord from the wall.
2. On the footboard control panel, hold down the bed motion lock button, the knee lock out button and the Reverse
Trend button. While still holding the buttons, plug the power cord into the wall socket. Release the footboard
buttons. Using the up/down arrow keys, scroll to the “Burn-In” item and hit “Enter.” Select “Burn-In Fowler” and hit
“Enter.” A message will appear: “Set Fowler to 0
0
”.
3. Using the footboard controls, run the Fowler to 0 degrees. Press the “Enter” button; a message will appear on the
display: “Set Fowler to 30 degrees.”
4. Using the footboard controls, run the fowler to 30 degrees. Press the “Enter” button; a message will appear on
the display: “Set Fowler to Upper Limit” (60 degrees).
5. Using the footboard controls, run the fowler to Upper Limit (60 degrees). Press the “Enter” button; a message will
appear on the display: “Fowler Calibration Complete” for two seconds, and then the user will be returned to the
“Burn-In” menu. The menu item should now read “Fowler Limits [SET]”.
6. Unplug the bed and plug back in to return to normal operating mode.
WARNING
Ensure there is no patient on the bed while the burn-in procedure is being performed.
Service Information
Summary of Contents for 3002
Page 2: ......
Page 82: ...Return To Table of Contents 82 3006 409 006 REV C www stryker com Bed Assembly Standard Bed ...
Page 83: ...Return To Table of Contents www stryker com 3006 409 006 REV C 83 Bed Assembly Standard Bed ...
Page 84: ...Return To Table of Contents 84 3006 409 006 REV C www stryker com Bed Assembly Standard Bed ...
Page 85: ...Return To Table of Contents www stryker com 3006 409 006 REV C 85 Bed Assembly Standard Bed ...
Page 91: ...Return To Table of Contents www stryker com 3006 409 006 REV C 91 Base Assembly Standard Bed ...
Page 92: ...Return To Table of Contents 92 3006 409 006 REV C www stryker com Base Assembly Standard Bed ...
Page 95: ...Return To Table of Contents www stryker com 3006 409 006 REV C 95 Base Assembly Standard Bed ...
Page 100: ...Return To Table of Contents 100 3006 409 006 REV C www stryker com Lift Assembly ...
Page 134: ...Return To Table of Contents 134 3006 409 006 REV C www stryker com Mechanical Litter Assembly ...
Page 135: ...Return To Table of Contents www stryker com 3006 409 006 REV C 135 Mechanical Litter Assembly ...
Page 164: ...Return To Table of Contents 164 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Page 165: ...Return To Table of Contents www stryker com 3006 409 006 REV C 165 Head End Siderail Assembly ...
Page 166: ...Return To Table of Contents 166 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Page 167: ...Return To Table of Contents www stryker com 3006 409 006 REV C 167 Head End Siderail Assembly ...
Page 168: ...Return To Table of Contents 168 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Page 189: ...Return To Table of Contents www stryker com 3006 409 006 REV C 189 Foot End Siderail Assembly ...
Page 190: ...Return To Table of Contents 190 3006 409 006 REV C www stryker com Foot End Siderail Assembly ...
Page 191: ...Return To Table of Contents www stryker com 3006 409 006 REV C 191 Foot End Siderail Assembly ...
Page 192: ...Return To Table of Contents 192 3006 409 006 REV C www stryker com Foot End Siderail Assembly ...
Page 193: ...Return To Table of Contents www stryker com 3006 409 006 REV C 193 Foot End Siderail Assembly ...
Page 200: ...Return To Table of Contents 200 3006 409 006 REV C www stryker com Notes ...
Page 244: ......