Return To Table of Contents
28
3006-409-006 REV C
www.stryker.com
Problem / Failure
Recommended Action
No power to bed.
A. Verify the power cord connections at the wall and the
bed.
B. Check the circuit breakers under the Litter/Fowler
section on the patient’s left side. If the circuit breaker
is tripped, reset it by pushing in.
C. Check for 230VAC at J1 on the power supply, Pin 1
and 2.
D. Check for DC voltages on J2 (Pins 1, 2, 3 & 6) on
power supply. See “Power Supply” in the Bed Circuit
Boards section for power supply voltage test points.
a. If voltage is present, check connector W on the
CPU board and check for the same DC voltages.
If OK, go to step E.
b. If voltage is not present, unplug connector W on
the CPU board and recheck for DC voltages at J2
on the power supply.
1.
If voltages come back, re-connect cable W to
the CPU board and go to step C.
2.
If DC voltage does not come back, replace
the power supply.
a. Unplug all connectors except for F, FF, O, and
W from the CPU board and recheck voltages on
connector W.
1.
If DC voltages come back, plug the cable
connections back in until problem comes
back, isolate the problem to a component or
assembly.
2.
If DC voltages do not come back, replace the
CPU board.
E. Check for 230VAC at connector O on the CPU board.
a. If voltage is present, replace CPU board.
F. If the CPU board has been replaced, reconfigure the
bed by resetting the: Bed Type, Bed Options, Serial
Number, Burn-in Limits (Fowler and Height) and
calibrating the scale.
G. Verify bed function and return to service.
Note
See the “Bed Circuit Boards” section for an outline of bed PCB’s and voltage test points.
Troubleshooting Guide
Summary of Contents for 3002
Page 2: ......
Page 82: ...Return To Table of Contents 82 3006 409 006 REV C www stryker com Bed Assembly Standard Bed ...
Page 83: ...Return To Table of Contents www stryker com 3006 409 006 REV C 83 Bed Assembly Standard Bed ...
Page 84: ...Return To Table of Contents 84 3006 409 006 REV C www stryker com Bed Assembly Standard Bed ...
Page 85: ...Return To Table of Contents www stryker com 3006 409 006 REV C 85 Bed Assembly Standard Bed ...
Page 91: ...Return To Table of Contents www stryker com 3006 409 006 REV C 91 Base Assembly Standard Bed ...
Page 92: ...Return To Table of Contents 92 3006 409 006 REV C www stryker com Base Assembly Standard Bed ...
Page 95: ...Return To Table of Contents www stryker com 3006 409 006 REV C 95 Base Assembly Standard Bed ...
Page 100: ...Return To Table of Contents 100 3006 409 006 REV C www stryker com Lift Assembly ...
Page 134: ...Return To Table of Contents 134 3006 409 006 REV C www stryker com Mechanical Litter Assembly ...
Page 135: ...Return To Table of Contents www stryker com 3006 409 006 REV C 135 Mechanical Litter Assembly ...
Page 164: ...Return To Table of Contents 164 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Page 165: ...Return To Table of Contents www stryker com 3006 409 006 REV C 165 Head End Siderail Assembly ...
Page 166: ...Return To Table of Contents 166 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Page 167: ...Return To Table of Contents www stryker com 3006 409 006 REV C 167 Head End Siderail Assembly ...
Page 168: ...Return To Table of Contents 168 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Page 189: ...Return To Table of Contents www stryker com 3006 409 006 REV C 189 Foot End Siderail Assembly ...
Page 190: ...Return To Table of Contents 190 3006 409 006 REV C www stryker com Foot End Siderail Assembly ...
Page 191: ...Return To Table of Contents www stryker com 3006 409 006 REV C 191 Foot End Siderail Assembly ...
Page 192: ...Return To Table of Contents 192 3006 409 006 REV C www stryker com Foot End Siderail Assembly ...
Page 193: ...Return To Table of Contents www stryker com 3006 409 006 REV C 193 Foot End Siderail Assembly ...
Page 200: ...Return To Table of Contents 200 3006 409 006 REV C www stryker com Notes ...
Page 244: ......