Menu Manager Configuration Errors
Error
Description
Recommended Action
140
scale frequency exceeds limits of
relay (1000 Hz)
.
150
scale frequency exceeds limits (10
kHz)
operation.
170
Output 1/2: Output Type
Configuration Error
This error is observed when the function type is 24V DC and the output type is changed from
Normally Open to Normally Closed. It is required for 24V DC output operation that the output type be
Normally Open.
171
Output 1/2: Output Type
Configuration Error
This error is observed when the function type is ADE and the output type is changed from Normally
Open to Normally Closed. It is required for ADE operation that the output type be Normally Open
190
Full Scale Flow: Entered Value
exceeds limits
Value entered exceeds the absolute maximum flow the meter supports. Reduce the value for this
parameter or consider increasing pipe diameter.
Display Error / Status Messages
Error Message
Possible Cause
Recommended Action
Err: Detector
.
Check detector and cable connections in accordance with this manual.
Contact Technical Support.
Supply voltage too low.
Contact Technical Support.
Grounded coils in meter.
Contact Technical Support.
Water in detector.
Contact Technical Support.
Err: Empty pipe
Pipe may not be full.
Make sure all trapped air is out of system.
.
Err: Full scale
.
.
Err: AD-Range
AD-Converter is exceeding signal limits.
Check the grounding scheme of the meter installation. See grounding section in this
manual. Verify pipe is not empty.
Err: AD-INT
Initialization of AD-Converter unsuccessful.
Contact Technical Support.
Err: Rollover
Rollover counters have exceeded limit.
Clear all totalizers.
Err: Rollover Status
Totalizer rollover has occurred.
Reload totalizer then clear all totalizers.
Err: Simulation
I/O simulator is enabled.
Disable simulator in I/O menu.