USING THE SSK
513
MAIN MENU PROGRAMMING OPTIONS
The following SSK
513
programming options are available from the
Main Menu
:
• Meter
Setup
• Measurements
• Inputs/Outputs
• Clear
Totals
• Communications
• Advanced
• Info/Help
• Language
Select
In the section that follows, the applicable security level for each menu option is indicated as follows:
A
S
U
Administrative
Service
User
Options that can be set at
Quick Setup
are indicated with:
The factory default values are shown, enclosed in brackets.
OTE:
N
Options that are listed [Factory Set] should not be changed without specific directions from authorized Badger Meter
personnel.
Meter Setup
Scale Factor
[0.0%]
Changing the scale factor lets you adjust the meter's accuracy without disturbing factory-set
parameters. You can tune the meter to meet changing application requirements. For example, if the
meter is under registering by 0.5 percent then set the scale factor to +0.5%. If the meter is over
registering by 0.5 percent then set the scale factor to –0.5%.
To set the Scale Factor, follow these steps from the
Meter Setup
menu:
1. Select
Scale Factor
, to open the
Detector Factor
display.
2. Set the Detector Factor value to the desired setting.
3. Press
[E]
, to save the new value and return to the
Meter Setup
menu.
Empty Pipe Detect
[Off]
When set to On, Empty Pipe Detect indicates to the outputs and the display that the meter is not
completely filled. When set to Off, empty pipe detect is disabled.
Enabling empty pipe detect requires a one-time calibration. Calibration is described in the
Advanced
menu section under Empty Pipe Cal.
To set Empty Pipe Detect, follow these steps from the
Meter Setup
menu:
1. Select
Empty Pipe Detect
to view the
Empty Pipe Detect
display.
2. Position the arrow next to
On
or
Off
.
3. Press
[E]
to save the Empty Pipe Detect On or Off and return to the
Meter Setup
menu.
Installation & Operation Manual
Page 29
March 2014