Wiring for Remote Configuration
44
40
44
46
44
45
13
12
11
shield (black)
pink
brown
shield (black)
white
shield (black)
yellow
green
green
yellow
shield (black)
white
shield (black)
brown
pink
shield (black)
E1
ES
E2
RS
EP
ES
CS C2 C1
Figure 17: Wiring for Remote Configuration
Remote style SSK
513
amplifier models can be ordered with standard cables measuring 15, 30, 50 and 100 feet. In addition,
cables up to 500 feet are available.
Junction Box
Amplifier
Connection No.
Description
Wire Color
Connection
11
Coil
Green
C1
12
Coil
Yellow
C2
13
Main Shield
Black (Red Ferrule)
CS
45
Electrode
White
E1
44*
Electrode Shield
Black
ES
46
Electrode
Brown
E2
40
Empty Pipe
Pink
EP
44*
Empty Pipe Shield
Black
ES
*Connections with the No. 44 are lying on the same potential.
Empty Pipe Detection Considerations
Take into account the following cable length and conductivity requirements, if you will be using empty pipe detection.
Cable Length (Feet)
Minimum Conductivity Required (μS/cm)
0*
5
100
20
500
100
* Meter Mount
Installation & Operation Manual
Page 17
March 2014