Inputs/Outputs
Digital Output
Preset Amount
[0.0]
S
Preset amount lets you set the reset value for the associated PS totalizer when the
digital input is set to Batch Reset.
To change the preset amount, follow these steps from the
Inputs/Outputs
menu:
1. Select
Digital Output 1
,
2, 3 or 4
and press
[E]
to open the
Digital Output
menu.
2. From the
Digital Output
menu select
Preset Amount
, and press
[E]
to open the
Preset Amount
display.
3. Enter the preset amount value. Press
[E]
to save the new parameter and return
to the
Digital Output
menu.
OTE:
N
You can only set one Preset Amount. If you set the Preset Amount for Digital
Output 1, it will be the same for 2, 3 and 4.
Set Point Min.
[0%]
S
This parameter establishes, as a percentage of full scale flow, the threshold at which
the output alarm will be activated. Flow rates below the threshold will activate the
output alarm.
To change the set point minimum, follow these steps from the
Inputs/Outputs
menu:
1. Select
Digital Output 1
,
2, 3 or 4
and press
[E]
to open the
Digital Output
menu.
2. From the
Digital Output
menu select
Set Point Min.
, and press
[E]
to open the
Set Point Min.
display.
3. Enter the set point minimum value. Press
[E]
to save the new parameter and
return to the
Digital Output
menu.
Set Point Max.
[100%]
S
This parameter establishes, as a percentage of full scale flow, the threshold at which
the output alarm will be activated. Flow rates above the threshold will activate the
output alarm.
To change the maximum set point, follow these steps from the
Inputs/Outputs
menu:
1. Select
Digital Output 1
,
2, 3 or 4
and press
[E]
to open the
Digital Output
menu.
2. From the
Digital Output
menu select
Set Point Max.
and press
[E]
to open the
Set Point Max.
display.
3. Enter the set point maximum value and press
[E]
to save the new parameter
and return to the
Digital Output
menu.
SSK
513
Electromagnetic Flow Meter
Page 36
March 2014