ADE
Protocol Type
The protocol type enables the ADE interface.
Selecting S1 or S2 will automatically configure
the Digital Input and Digital Output #1 for ADE
operation. Manually configuring the input and
output for ADE operation is not allowed and will
result in a on-screen error.
See "ADE Interface" on page 52 for further
details.
S1 – Standard ADE protocol
S2 – Enhanced ADE protocol, provides additional
digital information
Disabled – disables and removed ADE
configuration
Dial Type
4 – 8 dials
Analog Calibrate
Custom Settings
[Zero Scale: 0 mA]
[Full Scale: 0 mA]
S
To set the analog calibration custom settings, follow these steps from the
Advanced
menu:
1. Select
Analog Calibrate
to view the
Analog Calibrate
menu.
2. Select
Custom Settings
to view the
Custom Settings
display.
3. Select one of the following:
• Offset 4 mA
• Offset 20 mA
4. Configure desired offset.
5. Press
[E]
to save the option and to return to the
Custom Settings
menu.
6. Press
[E]
to return to the
Analog Calibrate
menu.
Factory Settings
[Factory Set]
To change the analog calibration factory settings, follow these steps from the
Advanced
menu:
1. Select
Analog Calibrate
to view the
Analog Calibrate
menu.
2. Select
Factory Settings
to view the
Factory Settings
display.
3. Select one of the following:
• Calibration
Point
A
• Calibration
Point
B
4. Set the calibration point to the measured output current.
5. Press
[E]
to save the option and to return to the
Factory Settings
menu.
6. Press
[E]
to return to the
Analog Calibrate
menu.
SSK
513
Electromagnetic Flow Meter
Page 44
March 2014
S