
12
Inside the circuit, the incoming pressurized medical gas, is sent to the mixing block by means of a device operated by a valve controlled by
the electronic circuit of the device. When the device is turned off the pressurized medical gas, even if introduced in the device, is not
delivered because intercepted by the electrovalve in closed position. Everything is controlled by the electronic circuit of the device, which
also controls the main breathing parameters.
This diagram is provided only to illustrate the operation of the device. It is not allowed to
intervene in any way on it as well as on the electronic circuit. If opening of the case of the device or any other type of unauthorized
modification will be found, the warranty would be voided and Spencer Italia will not have any liability related to the functionality and use
of the product.
9.
INSTALLATION AND START-UP
9.1
Installation
The installation of the device is a critical step to ensure a proper operation. When the product is recieved, verify that:
The packaging is intact and has no signs of impacts, falls and isn’t wet.
All the items in the accompanying list are present
The device does not show any kind of damage
Ensure that the vehicle in which you want to install the device, is manufactured in compliance with EN 1789.
The device, tested in the laboratory of a notified body, has successfully passed all the electromagnetic test required by the harmonized
standards. It is designed to be used in the electromagnetic environment described below. The customer, installer and user must be able to
ensure that these conditions are always respected.
Guide and Manufacturer's Declaration
190 ventilator is intended for use in the electromagnetic environment specified below. The customer or the user of the ventilator 190 must ensure that it is
used in such an environment.
EMISSION TESTS
Conformity
Guide to the electromagnetic environment
Emissions in RF CISPR 11
Group 1
The 190 ventilator uses RF energy only for its
internal functions. Its RF emissions are therefore
very low and unlikely to cause any interference
with electronic equipment nearby.
Emissions in RF CISPR 11
Class B
The 190 ventilator is suitable for use in all
environments including domestic as well as those
directly connected to a low-voltage public
network source supply of which supplies
buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2
Emissions as a result of voltage fluctuations /
flicker-IEC 6100 3-3
Compliant
Guide and Manufacturer's Declaration
The 190 ventilator is intended for use in the electromagnetic environment specified below. The customer and / or user of the ventilator 190 must ensure
that the device is used in such environment.
IMMUNITY TEST
Conformity level
Guida all’ambiente elettromagnetico
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8kV at contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV in air
I Floors should be wood, concrete or ceramic tile. If the
floors are covered with synthetic material, the relative
humidity should not exceed most 30%
Electrical fast transient/burst
IEC 61000-4-4
± 2kV power supply
± 1kV for input/ouput lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Over voltage
IEC 61000-4-5
± 0.5 , 1kV for line to line surge
0.5, 1, 2kV for line to ground surge
Mains power quality should be that of
a typical commercial or hospital
environment
Voltage dips
IEC 61000-4-11
0 % UT; 0,5 cycles at 0°,
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT 1 cycle and 70% UT 25/30 cycles (25 at 50Hz
and 30 at 60Hz)
Single phase ta 0°
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
lung ventilator 190 requires
continued operation during power
mains interruptions, it is recommended
that the lung ventilator 190
be powered from an uninterruptible
power supply or a battery.
Voltage interruptions
IEC 61000-4-11
0% UT; 250/300 cycles
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment
note UT is the value of the tension of power source