10
8.
TECHNICAL DATA AND COMPONENTS
Note: Spencer Italia S.r.l. reserves the right to make changes to specifications without prior notice.
8.1
Technical data
DIMENSIONALS
Width
330 mm
Height
210 mm
Depth
177 mm
Weight
3,5 ± 0,5 kg
Patient circuit weight
242 g
Patient circuit volume
400 ± 20ml
Power supply
Voltage
12 Vcc (-15% + 25%)
AC/DC Adapter
Input: 100-240VAC, 50/60Hz; 0,7-0,35A
Output: 12V DC
Power consumption max
700 mA
Main battery
12V NiMh 1,8Ah
Autonomy
About 5 hours
Charge time
About 12 hours
Secondary battery
9V 6LR61
Pneumatic supply
Intended gas
Medical oxygen
Input pressure
from 280 to 600 kPa
Maximum flow rate required
140 l/min
FUSE
F1L250V
Fast 5x20mm 1A
VENTILATION
Modes
CMV, TVM, A/CV
Volume/minute
from 2 to 20 L/min
Frequency (
F
)
from 5 to 40 BpM
Pressure limit
from 20 to 60 cmH
2
O
Tidal Volume
𝑉𝑇 =
𝑉𝑜𝑙𝑢𝑚𝑒
𝑚𝑖𝑛𝑢𝑡𝑒
𝐹
FiO
2
60% (AIR MIX) o 100% (NO AIR MIX)
I/E Ratio
1:2
High pressure alarm with frequency <20 bpm
60 mbar ± 5%
High pressure alarm with frequency >20 bpm
25 mbar ± 5%
Respiratory resistance
Valve
Filter
Breathing tube
2,5 cmH
2
O espirat.
1,73 cmH
2
O Inspirat.
at 50 l/min
2,5 cmH
2
O
a 60L/min
0,12 cmH
2
O at 30l/min
0,51 cmH
2
O at 60 l/min
Breathing tube compliance
Less than 2% of the volume
MANOVACUOMETER
Measuring range
from -20 to +80 mbar
Precision
1,6 (maximum error ± 1,3 mbar)
MAXIMUM DEVIATION FROM SETTINGS
Volume/minute
± 15%
Frequency
± 1 BpM
Pressure limit
± 5 mbar
Allarms
High priority
High pressure – Low pressure – Apnea – Pneumatic supply –
Low battery
Maximum audio paused time
100 s
Acoustic level
80 ± 10db
Classifications
Classification according to directive 93/42/CE
IIb
Classification according to IEC 60601-1
Class II if externally powered
Internally powered if operating with internal battery
Applied part (Breathing tube, filter, connector, valve, mask)
BF
Enclosure protection according to IEC 60529
IP24
Protected against ingress of particles with diameter greater
than 12mm
Protected against splashing water from all directions