Trigger modes and appendant registers
Channel Trigger
(c) Spectrum Instrumentation GmbH
115
Detailed description of the channel trigger modes
For all channel trigger modes, the OR mask must contain the corresponding input channels (channel 0 taken as example here):.
Channel trigger on positive edge
The analog input is continuously sampled with the selected
sample rate. If the programmed trigger level is crossed by
the channel’s signal from lower values to higher values (ris
-
ing edge) then the triggerevent will be detected.
These edge triggered channel trigger modes correspond to
the trigger possibilities of usual oscilloscopes.
Channel trigger on negative edge
The analog input is continuously sampled with the selected
sample rate. If the programmed trigger level is crossed by
the channel’s signal from higher values to lower values (fall
-
ing edge) then the triggerevent will be detected.
These edge triggered channel trigger modes correspond to
the trigger possibilities of usual oscilloscopes.
Channel trigger on positive and negative edge
The analog input is continuously sampled with the selected
sample rate. If the programmed trigger level is crossed by
the channel’s signal (either rising or falling edge) the trig
-
gerevent will be detected.
These edge triggered channel trigger modes correspond to
the trigger possibilities of usual oscilloscopes.
Register
Value
Direction
Description
SPC_TRIG_CH_ORMASK0
40460
read/write
Defines the OR mask for the channel trigger sources.
SPC_TMASK0_CH0
1h
Enables channel0 input for the channel OR mask
Register
Value
Direction
set to
Value
SPC_TRIG_CH0_MODE
40610
read/write
SPC_TM_POS
1h
SPC_TRIG_CH0_LEVEL0
42200
read/write
Set it to the desired trigger level relatively to the channel’s input range.
board dependant
Register
Value
Direction
set to
Value
SPC_TRIG_CH0_MODE
40610
read/write
SPC_TM_NEG
2h
SPC_TRIG_CH0_LEVEL0
42200
read/write
Set it to the desired trigger level relatively to the channel’s input range.
board dependant
Register
Value
Direction
set to
Value
SPC_TRIG_CH0_MODE
40610
read/write
SPC_TM_BOTH
4h
SPC_TRIG_CH0_LEVEL0
42200
read/write
Set it to the desired trigger level relatively to the channel’s input range.
board dependant