104
(c) Spectrum Instrumentation GmbH
Main analog external trigger (Ext0)
Trigger modes and appendant registers
The following pages explain the available modes in detail.
Trigger on positive edge
The trigger input is continuously sampled with the selected
sample rate. If the programmed trigger level is crossed by
the trigger signal from lower values to higher values (rising
edge) then the trigger event will be detected.
This edge triggered external trigger mode correspond to
the trigger possibilities of usual oscilloscopes.
Trigger on negative edge
The trigger input is continuously sampled with the selected
sample rate. If the programmed trigger level is crossed by
the trigger signal from higher values to lower values (falling
edge) then the trigger event will be detected.
This edge triggered external trigger mode correspond to
the trigger possibilities of usual oscilloscopes.
SPC_TMASK_EXT0
2h
Enables the main external (analog) trigger 0 for the mask.
Register
Value
Direction
set to
Value
SPC_TRIG_EXT0_MODE
40510
read/write
SPC_TM_POS
1h
SPC_TRIG_EXT0_LEVEL0
42320
read/write
Set it to the desired trigger level in mV
mV
Register
Value
Direction
set to
Value
SPC_TRIG_EXT0_MODE
40510
read/write
SPC_TM_NEG
2h
SPC_TRIG_EXT0_LEVEL0
42320
read/write
Set it to the desired trigger level in mV
mV
Summary of Contents for M2p.59 Series