MiniCapt® Mobile Operations Manual
8-1
Chapter 8
Warnings and Alarms Setup
This chapter describes configuration of warnings and alarms for the MiniCapt Mobile
Microbial Air Sampler.
Warnings:
A warning will occur if the flow is ± 5% of the calibrated flowrate (up to ±20%).
The warning will be visual on the screen as well as audible if sound is configured.
See
Alarms:
An alarm will occur if the flow is outside of 20% of the calibrated flowrate.
The alarm will be visual on the screen as well as audible if selected and the MiniCapt
will turn off.
An alarm comment will need to be entered using the drop down menu.
See
View the Warnings and Alarms List Screen