Installation
Chapter 2
: Unpacking and Installation
MiniCapt® Mobile Operations Manual
2-9
Adjust the Plate Holder Diameter
If the agar plate you intend to use does not fit inside the plate holder brackets, you can
adjust the brackets in the plate holder to increase or decrease its diameter. The
MiniCapt Mobile can be used with an agar plate having an external diameter between
86 and 92 mm.
This procedure requires:
•
A 1.5 mm hex driver
•
An agar plate
WARNING
Remove power to the unit prior to removing sampler head.
>> To adjust for the plate’s diameter:
1.
Remove the sampler head (see
WARNING
After removing the sampling head, take care to ensure loose items
are not dropped into the opening to the blower.