Configuration
Chapter 1
: Introduction
MiniCapt® Mobile Operations Manual
1-7
Configuration
The MiniCapt Mobile requires configuration of the following operating parameters for
sampling:
•
Choice of sampling mode (Time or Volume)
•
(Time mode) Sampling duration
•
25 LPM: from 1 to 400 minutes
•
50 LPM: from 1 to 200 minutes
•
100 LPM: from 1 to 100 minutes
•
(Volume mode) Sampling volume in liters from 1 to 9999 L
•
Start delay: time delay before sample is started 0 to 139 minutes
•
Number of sampling intervals: from 1 to 1000 intervals
•
Waiting (pause) between an interval and the next, from 0 to 139 minutes
Available Memory
The MiniCapt Mobile Microbial Air Sampler is outfitted with non-volatile EEPROM
memory that is able to store up to:
•
3,000 samples
•
400 areas/locations
•
50 recipes
NOTE:
If there are 3,000 records in memory, the oldest record will be automatically
discarded for storage of the new record.