Operation
When power failure occurs
Thermo Scientific
WX Ultra Series
2-67
Figure 2-16
Removing the front cover
(4) Open the air vent (by turning the vacuum release screw on the left of the vacuum chamber
counterclockwise) to let air into the rotor chamber. When the pressure in the rotor chamber reaches
atmospheric pressure, do not forget to tighten the vacuum release screw as it was before. (See Fig.
2-16.)
(5) Pull the door unlock wire on the right of the vacuum chamber and, at the same time, push the door
handle. Then the door opens. When opening the door, confirm that the rotor is not rotating. If it is
still rotating, immediately close the door.
(6) Remove the rotor. After the removal, close the air vent and put the front cover back to the
ultracentrifuge. Insert the top surface of the front cover in the bottom surface of the front side of the
table and put the bottom surface of the front cover on the support plate of the front cover by reversing
the removing procedure. Attach it on the frame with four M5 hexagon head bolts.
Table
Front cover
Cabinet cover
M5 hexagon head
bolt for fixing the
front cover (4 pcs.)
WARNING
Never touch the rotor while it is rotating.