Operation
Rotor management
Thermo Scientific
WX Ultra Series
2-65
Updating the existing data for a rotor
When a rotor is registered in the Sorvall WX Ultra series ultracentrifuges without a Rotor Management
function, both Total number of runs and Total number of hours can be updated. The updated value is
accepted only when it is greater than the registered value. It must not be smaller than the registered
value.
To update the existing life management data for a standard rotor, perform the following procedure:
Step
Key operation
Screen display and considerations
1
While in the Run Screen, use the cur-
sor keys to move the cursor to RLM
and press ENTER key.
2
Use cursor keys to move the cursor to
the T-890 line to be updated and press
the ENTER key.
3
Select “Creating and Changing” by
pressing the following keys.
4
Enter the updated total number of runs
and the total number of hours by using
numeric keys.
Total number of runs
500.0 ---> 505.0
Total number of hours
1200.4 ---> 1210.4
• The display switches to the
Rotor Management screen.
• In the Runs field, the cursor
waits for an input.