Operation
How to use the FUNCTION field
2-30
WX Ultra Series
Thermo Scientific
2.
How to perform a programmed operation
Shown below is how to perform a “programmed operation”, that is, how to recall a stored set of run
conditions and run this centrifuge accordingly.
(a) If you know which memory unit number you need
Step
Key operation
Screen display and considerations
1
Turn on the POWER switch on the cen-
trifuge.
>> The panel display appears.
>> The door unlocks.
2
Install a rotor.
>> Install the rotor securely on the shaft.
3
While in the Run Screen, use cursor
keys to move the cursor to PROG • RTC
and then press ENTER key.
Next, move the cursor to PROG and
press ENTER key.
4
Enter the memory number you wish to
operate.
Example: to call memory number 3,
press
>> The message field in the bottom row displays PROG.
>> If the STEP field displays what is shown below, this means that the memory
unit you have just called stores multiple run conditions (stepmode feature). For
details, see Section
Step-mode operation
.
>> The system only accepts numbers 1-20.
5
Run the centrifuge under normal oper-
ation without making changes to the
run conditions.
>> Run this machine according to
Normal operation
.
>> If you make changes to the run conditions (such as SPEED and TIME) after call-
ing a program, the program you have just called is canceled. You must call it
again to use it.
• The FUNCTION field switches
to the PROGRAM field.
• The MEMORY No. setting field
blinks.
• The run conditions in memory
appear in the respective fields.
In the Run Screen, the run conditions of
this step are displayed
(the first step
in this case).