Operation
How to use the FUNCTION field
Thermo Scientific
WX Ultra Series
2-55
4
To select a desired rotor from the
“Rotor Management”:
(1) Select “Rotor Management”.
(2) Select a desired rotor using the
upward or downward cursor key
.
(When using fixed angle rotor T-8100)
(3) Select “Yes”.
>> The RCF value corresponding to the currently set speed is displayed in the RCF
setting field.
(--- --- --- --- --- ---: Displayed at the initial time of no setting.)
5
Use cursor keys to move the cursor to
the RCF setting field, and enter a
desired centrifugal force.
Example: RCFmaX = 5000 x 100g
• Specify the RCF value to hundreds.
If you make no input for 30 seconds, the cursor will disappear, and you will be
unable to make an input. At that time, press a cursor key.
• The display switches to the
Rotor Management screen and
the Rotor List is displayed.
• The display switches back to
the Run Screen and the selected
rotor type is displayed in the
message display field of the
bottom row.
RCF display field
RCF setting field
• The value you have just entered
appears in the RCF setting
field.
• Based on the rotor and RCF
value, the system automatically
calculates the speed and
displays it in the SPEED
display field.