Operation
How to use the FUNCTION field
Thermo Scientific
WX Ultra Series
2-41
4. Control temperature: 4 °C
5. Acceleration mode: 9
6. Deceleration mode: 7
Figure 2-14
A typical RTC run
In this example, you set the above run conditions (2) through (6), set the finish time for RTC run to
9:30 a.m., October 1 and start the centrifuge.
(You can make an identical setting by setting the start time to 5:30 a.m. instead of setting the finish
time to 9:30 a.m.)
100,000
Sp
ee
d (rpm
)
START key ON
4:30 p.m, October 1
Automatic start at 5:30 a.m., October 2
4 h
Stop around 9:34 a.m.
Finish time