Operation
How to use the FUNCTION field
2-44
WX Ultra Series
Thermo Scientific
1. For an RTC run, the run mode indicator on the display panel displays the following:
Figure 2-15
Run mode display (RTC)
2. You cannot make an RTC setting in any of the following cases:
a. When the Run Screen is set to HOLD (continuous run)
Set the run time (centrifugation time) not to HOLD but to a numerical value.
b. When it is past the start time
Set the start time to a time later than the current time.
c.
When the start time is more than 20 days after the current time
Set the time to a time no more than 20 days afterwards.
3. To change the run time (centrifugation time) after making an RTC setting, cancel RTC and then
set a new run time.
4. To perform a combination of a programmed operation (including a step-mode operation) with an
RTC run, call a program memory number, then set RTC.
The system calculates the total run times of all steps of the programmed operation and calculates
the start time for RTC.
Therefore, you cannot recall the program memory number after setting RTC.
5. To stop this ultracentrifuge in RTC operation, press STOP key. The system then stops RTC and
stops the rotor.