Operation
Basic operating procedure
Thermo Scientific
WX Ultra Series
2-11
Basic operating procedure
There are two basic modes of operation, normal and zonal. The procedures for these two modes are shown in Fig. 2-5.
START
Turn on the POWER.
Open the chamber door and
install the rotor.
Normal run
Close the door and set the run
conditions.
Press VACUUM.
For information on how to set
the run conditions, see
Section
Basic operation
Vacuum pump is ON.
Temperature control starts.
Check the set run conditions
for accuracy.
Press START.
Rotor accelerates to the set speed, and then
high-speed centrifugation begins.
: Indicates panel key.
Zonal run
Reaches the set time,
or press STOP.
Rotor comes to a stop.
Press VACUUM.
Temperature control is OFF.
Vacuum pump is OFF.
Air valve is open and door lock
is released.
Remove the rotor and turn off
the POWER switch.
END