Structure and function
IMPORTANT:
SKF will not accept liability for any damage caused if
the telescopic pillar CPMT is not used with a suitable operating
device and supporting metal plates on outer and inner tube base
plates.
4.5 Operating elements
Requirements for operating elements (mandatory in medical
applications):
General safety requirements according to IEC 60601-1 currently
valid edition, harmonized with national directives (Means of protection:
1 Means Of Patient Protection [MOPP]).
Suitable SKF operating elements are listed in the section
Accessories
.
4.6 Options
Options can be recognized from the type designation on the type
plate.
Voltage
The option voltage defines the use for 24-30 V DC.
Load
The CPMT is available in two standard load options, 5000 N or
6000 N push force. Pull load is 4000 N for both options.
Retracted length
Two standard retracted length options are available, defined by the
relationship to the stroke of the pillar. The options to choose from
are ‘stroke/2 + 120 mm’ or ‘stroke/2 + 240 mm’.
Stroke length
The available stroke length for CPMT is from 300 mm to 600 mm
in 10 mm steps for retracted length ‘stroke/2 + 240 mm’, and
from 400 mm to 600 mm in 10 mm steps for retracted length
‘stroke/2 + 120 mm’
Electrical options
The CPMT pillar comes equipped as standard option with a 2-hall
encoder and end limit switches.
Cable options
Available cable options are: With DIN-8 cable, in lengths of 1m or
2.3m; or with flying leads termination, in length of 2.3m.