PVA-3000 Reference Manual
December 2, 2019
Sifos Technologies
Page
46
The “nominal” transmit level is affected by the 4-pair Coupler in Figure 2.1 as described in Section 2.1. When
Transmit Level
is set to
Max
, the output level is increased by about 1 dB from that nominal level and when set to
Min
, it is reduced by about 1 dB (
see Section 2.2.11 for further detail
). Transmit Level configuration changes are only
applied when the
Apply Level
button is pressed.
Besides
Automatic
and
Nominal
,
Transmit Slew
may be configured to
Fast
and
Slow
.
Nominal
would generally
represent an edge transition of 4 nsec per 0.8V transition, or a slew rate of 0.2V/nsec. The
Fast
slew setting would
result in an edge transition of ~3 nsec per 0.8V, or 0.27V/nsec slew rate, while the
Slow
slew setting will configure
edge transition to approximately 4.5 nsec per 0.8V, or a .18V/nsec slew rate. Transmit Slew configuration changes are
only applied when the
Apply Slew
button is pressed.
3.7.
Configuration Replication
Five of the tabbed menus in PVA Interactive support various forms of resource configuration and therefore include a
capability to copy, or replicate, resource configurations to other test ports. Those menus are
Port Config
,
Link
Config
,
Link Monitor
,
Meters
, and
Packets
. Each of these menus includes a
Copy Settings
button that will open the
dialog shown in Figure 3.13. In the
Port Config
and
Link Config
tab menus, the settings to be copied will be the
selected subset of all configurations controlled by those
menus.
This dialog indicates the current
Copy from
port, that is, the
port from which settings are to be copied. The menu also
includes enabled checkboxes for every other available PVA
test port in the currently connected instrument. To copy
settings to particular PVA test ports, the slot-port check boxes
for those ports are selected and the
DONE
button in pressed.
Configuration settings can be copied from the current test port
to all other PVA test ports simply by selecting the
ALL
PORT
S checkbox and pressing
DONE
. If no destination ports are selected, then no configuration settings will be
copied when the menu is closed.
Finally, it should be noted that if
Copy Settings
is selected before a particular setting has been applied (e.g. an
Apply
Settings
button is annunciating a pending setting change or a meter measurement is not yet initiated after selecting a
meter configuration), that generally the
Copy Settings
button will first apply pending settings to the current test port
prior to copying those test port settings to other test ports.
3.8.
Link Configuration Menu
The remaining tab menus in PVA Interactive all perform operations that involve the TEST PHY. Therefore, they are
only relevant when the TEST
PHY is connected to the
TEST
interface via the
Port
Connection
in the Port Config
menu.
Link Configuration (
Link
Config
) is the second tab menu
within the PVA Interactive
Main Menu (see Figure 3.14).
This menu provides control of
Ethernet link modes, access to
test signals, and control of the
TEST PHY receiver. It
consists of five sub-menus:
Link Config
Gigabit Config
Test Signal
Test PHY
(Selected)
Actions
Figure 3.13
Configuration Replication Menu
Figure 3.14
Link Configuration Menu
Summary of Contents for PhyView PVA-3000
Page 22: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 22 ...
Page 70: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 70 ...
Page 108: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 108 ...
Page 124: ...PVA 3000 Reference Manual December 2 2019 Sifos Technologies Page 124 ...