Details of last calibration
This menu shows the details of the last manually performed calibration.
The following is displayed:
•
Time stamp of the last calibration
•
Measured value before the calibration
•
Measured value after the calibration
•
Drift of last calibration (difference between setpoint value and actual value,
calculated by the device)
•
Total deviation for zero point
•
Total deviation for calibration gas
The drift values are calculated by the respective device itself. The values before and
after the calibration are read by the system software shortly before and after the
calibration. This may result in slight differences between the drift and the values before
and after calibration.