Technical specifications
Set CEM CERT
Operating Instructions, 06/2021, A5E34491309-12
109
Physical setting data
Gas cooler
Se4
°C, set at factory
Monitoring of heated sample gas line -EH1
•
K2 - Monitoring
•
K1 - Control
Setting on -KT1
T
min
+140
°C
T
set
+180
℃
1)
Temperature setting on probe
T
set
+180
℃
1)
Cabinet temperature monitoring -KT2
Tmax +35
°C
Cabinet air conditioning
(manufacturer's setting on air conditioning
unit)
15
°C ... 35
°C
Thermostat for frost protection heater -EH10
8
°C
1)
The setpoint temperature of the heated sample gas line and the probe may differ depending on the acid dew point
of the gas. The maximum setpoint temperature depends on the type of probe and the type of sample gas line.
Observe the information provided by the respective manufacturer.
Time intervals
Calibration
•
ULTRAMAT 23
•
ULTRAMAT 6
24 h must be set for the automatic zero point calibration (Thermo
AUTOCAL function activated).
A weekly interval must be provided for the automatic zero and
reference point calibration.
Maintenance
See current certificate