Recalibration
7.4 Recalibration
BW500 and BW500/L
80
Operating Instructions, 12/2016, A5E33482052-AD
Span Calibration. Current Span
Setup test. Press ENTER to start
41900
For example: the current span count.
Note
End of span calibration. Remove the test weight and return to RUN.
7.4.7
Direct Span
Direct span entry (P368) is intended for use when replacing software or hardware, and when it is
not convenient to perform an initial span. A record of the last valid span count is required.
Access P368 and enter EDIT mode
P368-01 Direct Span Entry
Enter Span Count
E
0
Press the following keys in a given sequence:
P368-01 Direct Span Entry
Enter Span Count
V
41900
For example: the last valid span count.