Installation
5.4 Interconnection
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
31
Note
Color code for load cells used on Siemens weighfeeders may be different than shown.
Please refer to the weighfeeder wiring diagram.
Where separation between the BW500 and BW500/L and belt scale exceeds 150 m (500 ft.), or
legal for trade certification:
1.
remove jumpers from BW500 and BW500/L terminal 11/12 and 13/14.
2.
run additional conductors from:
BW500 and BW500/L terminal 12 to scale 'red'
BW500 and BW500/L terminal 13 to scale 'blk'
If the load cell wiring colors vary from those shown, or if extra wires are provided, consult
Siemens.
Summary of Contents for Milltronics BW500
Page 1: ...Operating Instructions Milltronics BW500 and BW500 L 12 2016 Edition Integrators ...
Page 12: ...Table of contents BW500 and BW500 L 10 Operating Instructions 12 2016 A5E33482052 AD ...
Page 14: ...Safety Notes BW500 and BW500 L 12 Operating Instructions 12 2016 A5E33482052 AD ...
Page 16: ...The Manual BW500 and BW500 L 14 Operating Instructions 12 2016 A5E33482052 AD ...
Page 48: ......
Page 154: ......
Page 218: ...Index BW500 and BW500 L 216 Operating Instructions 12 2016 A5E33482052 AD ...
Page 219: ......