Start Up
6.4 Load Cell Balancing
BW500 and BW500/L
64
Operating Instructions, 12/2016, A5E33482052-AD
Note
The moisture meter is ignored during calibration. If inclinometer is used, then calibration is
adjusted based on incline angle.
Span Calibration
When performing a Span Calibration where the test reference is ECal (P002 = 3), the supplied
test weight or test chain must not be applied, and the conveyor must be run empty.
Note
To obtain an accurate and successful calibration, ensure that the required criteria are met.
Refer to Calibration Criteria (Page 199).
With the conveyor stopped and locked out, apply the test weight or chain to the scale as
instructed in the associate manuals; then start the conveyor.
Press the following key:
Span Calibration Current Span
Setup test Press ENTER to Start
0
For example: the current span count.
Press the following key:
Initial Span Calibration in progress
Current Reading
%
####
The span count being calculated while calibration is in progress.
The duration of the Span calibration is dependent on speed (P014), length (P016), and
revolutions (P360) of the belt.
Span Count Too Low
Press CLEAR to continue
●
signal from load cell too low, ensure proper test weight or chain is applied during calibration
●
check for proper load cell wiring and ensure shipping brackets have been removed.
Summary of Contents for Milltronics BW500
Page 1: ...Operating Instructions Milltronics BW500 and BW500 L 12 2016 Edition Integrators ...
Page 12: ...Table of contents BW500 and BW500 L 10 Operating Instructions 12 2016 A5E33482052 AD ...
Page 14: ...Safety Notes BW500 and BW500 L 12 Operating Instructions 12 2016 A5E33482052 AD ...
Page 16: ...The Manual BW500 and BW500 L 14 Operating Instructions 12 2016 A5E33482052 AD ...
Page 48: ......
Page 154: ......
Page 218: ...Index BW500 and BW500 L 216 Operating Instructions 12 2016 A5E33482052 AD ...
Page 219: ......