
Image Quality
9 - 13
Siemens AG
SPB7-250.815.02
Page 13 of 20
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
If you
Reject
the image, a new image can be acquired for the Gain Calibration.
Correct the problem select the same image from the
Procedure: Calibration
Procedure
and continue with step 12.
14.
Wait approximately 10 seconds for the image to be calculated (observe the status
bar) and then double click on the next image in the
Procedure: Calibration
Procedure
and proceed with step 12 until you have a total of 8 images.
15.
Click
Done
to exit the Calibration.
16.
Close the examination.
You would "Reject" an image if it exhibits any edge cut-off due to
collimation or misalignment or if there are any artifacts from debris
or obstructions.
NOTE