MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 6 of 10
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
8 - 6
Checking dose settings
-
dose value
in the table above. The measured AEC dose values shall be
between:
H -> 120
µ
Gy to 140
µ
Gy
D -> 180
µ
Gy to 200
µ
Gy
- image
pixel mean value
Display the image with the
syngo Image Viewer
. To do so click on the
Viewing
tabcard, load the image with the
Patient -> Browser
and select
Tools -> Rect-
angle
.
Draw an area of approximately
1 sq.cm
and measure the pixel mean
value. You have to measure approximately
2 cm
above the chest wall at the
center of the detector.
Criteria: The expected pixel mean values are:
H = 250 ..... 350
D = 300 ..... 400
8.
If the
H & D
dose settings have to be adjusted, proceed with Chapter 8 "Adjust the
H and D Settings" and repeat the H & D dose check for Mo/Mo again.
9.
Record the measured values in the
"