MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 6 of 6
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
3 - 6
Power up the MAMMOMAT Stand
Startup of the Acquisition Workstation
3
System specific data has to be entered and set.
The startup and configuration of the Acquisition Workstation is described in the document
Startup, WH AWS
(SPB7-250.815.02...).
Arranging the swivel arm
3
1.
Remove the protective strips from the metal curtain.
2.
Run the lifting carriage upwards.
3.
Attach the protective strips back onto the metal curtain.
4.
Rotate the swivel arm system to
0 degrees
by pressing one of the switches for
clockwise rotation on the supporting arm.
5.
System and mains
OFF
.
Never rotate the swivel arm with the detector wing in the parking
position. This will damage the main cable harness.
WARNING