Collimator Adjustment
7 - 5
Siemens AG
SPB7-250.815.02
Page 5 of 6
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
10.
Continue with Step 7 until the beam field and light field are identical.
11.
Save
the calibration and
Exit
the Service Software.
12.
Record the measured values in the
Collimator Wing Difference Calibration
7
The collimator wing difference calibration has to be set to
0
.
1.
Start the service software on the Service PC and select:
Configuration -> Collimator -> Calibration of wing differance
2.
Save
the calibration and
Exit
the Service Software.
"
Fig. 4
Collimator wing difference calibration