Startup Protocol
11 - 7
Siemens AG
SPB7-250.815.02
Page 7 of 10
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
OPDOSE Settings
11
Results from the chapter "Checking OPDOSE" on Page 8 - 3.
Checking Grid Lines
11
Results from the chapter "Checking Grid Lines" on Page 9 - 1.
Program
button
Thickness
kV
Anode/filter
combination
AEC
1
2
3
4
Tab. 6
OPDOSE Settings
Expo-
sure
Exp Mode mAs
PMMA /
mm
Tube
Angle
visible
grid lines
(yes/no)
X
Y
Mean pixel
value differ-
ence [%]
1
mAs
32
30
0
o
2
mAs
56
40
0
o
3
mAs
200
60
0
o
4
AEC "H"
-
40
0
o
5
AEC "H"
-
40
+ 60
o
6
AEC "H"
-
40
- 60
o
Tab. 7
Measuring Grid Lines