MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 2 of 20
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
9 - 2
Image Quality
2.
Select the Procedure
User 1
from the examination tab card.
3.
Select the exposure parameters:
28 kV
and
Mo/Mo.
4.
Make sure that a 1:1 view is selected
,
to have a better view.
Select
Image -> Change Layout -> a 1:1
5.
Do all exposures according Tab. 1.
6.
Activate magnification
(Image -> Magnify by 2.0)
and look for grid lines over the
plexi phantom area.
7.
Test 1 Protocol
Remember, you can always add or remove images to
the "Procedure" by using the right mouse click.
Exposure Exp Mode mAs
PMMA /
mm
Tube Angle
Visible grid lines
(yes/no)
1
mAs
32
30
0
o
2
mAs
56
40
0
o
3
mAs
110
50
0
o
4
AEC "H"
-
40
0
o
5
AEC "H"
-
40
+ 60
o
6
AEC "H"
-
40
- 60
o
Tab. 1
Checking visibility of grid lines
Fig. 2
Select Procedure for grid line test
NOTE