Startup Protocol
11 - 5
Siemens AG
SPB7-250.815.02
Page 5 of 10
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
30 kV
10 mAs
W/Rh
_____V x 5 kV =
_____V x ( _____s) x 40 mA =
______ kV
______ mAs
25 kV
-
Mo/Mo
_____V x 5 kV =
_____V x ( _____s) x 40 mA =
______ kV
______ mAs
Display ______ mAs
25 kV
-
W/Rh
_____V x 5 kV =
_____V x ( _____s) x 40 mA =
______ kV
______ mAs
Display ______ mAs
kV
mAs
Anode/
filter
Measured values
Result
kV must be
&
5%
mAs must be
&
10%
Tab. 2
Tube high voltage, current and mAs values