Mechanical Checks
5 - 3
Siemens AG
SPB7-250.815.02
Page 3 of 12
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
Stand covers
5
1.
The covers are delivered with a protective plastic film, which shall be removed and
discarded.
2.
Make sure that clips are positioned over the holes where the screws (1/Fig. 3) are
to be fastened. These clips enable the screws to fasten the covers.
3.
Fit left and right front covers (2/Fig. 3) to the stand and fasten with screws and
contact washers (3/Fig. 3).
4.
The edges of the metal curtain of the stand are very sharp.
They may cause severe injury.
Remove the protective strips carefully when the covers are to be
mounted. Store the strips in the holders provided on both sides of
the curtain.
When mounting the stand covers, make sure that the screws at
the top of the stand are securely tightened and have functioning
contact washers.
These screws are used to establish protective ground connec-
tions.
Fig. 3
Mounting the stand covers on a stand with single rear cover.
Be careful not to damage the cables at the cable outlet!
WARNING
NOTE
MAM00214
1
2
3
4
5
7
6
8
NOTE