Detector installation
4 - 5
Siemens AG
SPB7-250.815.02
Page 5 of 6
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
Mounting the grid
4
The materials needed for the grid installation can be found in the accessory material box.
1.
Insert the left and right grid slider (1/Fig. 7) from the back into the rail.
2.
Insert the left and right grid holder into the grid slider (2/Fig. 7).
3.
Line up the grid to the front of the table, the front switch has to be engaged. Screw
the grid (3/Fig. 7) to the grid holder (4/Fig. 7). The marking has to be on top, text
must be readable.
4.
Connect the cables (5/Fig. 7) for the magnification table, according to the labels,
on both sides.
Fig. 7
Grid installation
Make sure that the cables are connected according to the labels,
otherwise the D805 will burn up.
S2
S3
1
2
3
4
5
WARNING