MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 2 of 6
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
7 - 2
Collimator Adjustment
•
Moving the collimator
IN
is done by
DECREASING
the
YC
value and
INCREASING
the
YS
,
XL
, and
XR
values.
Collimator check
7
You have to perform the check for all foci and Mo/Mo and W/Rh anode/filter combinations.
1.
Make sure that a patient is registered.
2.
Mount the Collimator Mounted Phantom.
3.
Connect the compression simulator.
4.
Selecting
25 kV
and
50 mA
s on the control console.
5.
Start an exposure.
6.
View the image with a magnification of 4.
7.
Slide the magnification glass along the image edges of the image and check if
white lines are visible.
Collimator beam calibration
7
Collimator calibration is carried out for each focus including the field light focus.
1.
Connect the compression plate simulator and the collimator mounted plexi phan-
tom.
If you do not see white lines at the edges of the image, the colli-
mator beam calibration is correct and you can continue with the
"Collimator field light calibration" on Page 7 - 4.
Otherwise continue with the "Collimator beam calibration" on
Page 7 - 2