
MAMMOMAT
Novation
DR
SPB7-250.815.02
Page 18 of 20
Siemens AG
System
Rev. 02
09.04
CS SD 24
Medical Solutions
9 - 18
Image Quality
7.
8.
Set the following exposure parameter on the control console of the MAMMOMAT
Novation
DR
:
Mo/Mo, 28KV, 250mAs
9.
Insert the collimator mounted plexi phantom.
10.
Mount the compression plate simulator.
11.
Double click on the
first
Image of the 8 at the
Procedure: Calibration Procedure
12.
Start an exposure.
Wait until the image is loaded into the viewer. The following lines are displayed in
the message bar.
Image acquisition is in progress .....
Saving processed Image .....
Loading Image ......
13.
Click the
Accept
or
Reject
button.
If you want to end the Bad Pixel Mapping operation at this point,
click the "Cancel" button.
Fig. 19 Acquisition System Menu
NOTE