Image Quality
9 - 19
Siemens AG
SPB7-250.815.02
Page 19 of 20
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
If you
Accept
the image, the new pixel map is created. No further images have
to be taken.
If you
Reject
the image, the Bad Pixel Mapping menu is displays. Correct the
problem and return to step 12.
14.
Click
Done
to exit the Bad Pixel Mapping.
15.
Close the examination.
If you want to end the Bad Pixel Mapping operation at this point,
click the "Cancel" button.
You would "Reject" an image if it exhibits any edge cut-off due to
collimation or misalignment or if there are any artifacts from debris
or obstructions.
Fig. 20 Bad Pixel Mapping - Accept Image
NOTE
NOTE