Final procedures
10 - 3
Siemens AG
SPB7-250.815.02
Page 3 of 4
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
Service PC and measuring instruments
10
Saving the configuration file
10
Main menu:
1.
Select
Configuration
Save config file
.
2.
Press <F2> to save data.
Reset the exposure counter
10
Main menu:
1.
Select
Service
Reset exposure counter
.
2.
Delete the error memory with <Y>.
Deleting the error buffer
10
Main menu:
1.
Select
Service
Delete error buffer
.
2.
Delete the error memory with <Y>.
Saving the MAMMOMAT Parameters
10
Main menu:
1.
Select
Backup
Copy installation area to floppy
All
.
2.
Leave the program with <F10> when the backup is ready.
Remove the measuring instruments
10
1.
Turn
OFF
the MAMMOMAT.
2.
Remove the service PC.
3.
Remove connected measuring instruments.
Checking the protective grounding resistance
10
Using the protective ground wire tester, measure the resistance between the protective
ground terminal in the stand (Mains input terminal) and all metallic parts of the equipment
which the patient/operator may come in contact with, incl. all object tables.
The protective ground resistance must not exceed
0.1
Ω
.
Make sure that the backup floppy is inserted.
Keep the backup floppy with other documentation for the
MAMMOMAT
Novation
DR
in question.
NOTE
NOTE